Optional Configuration
Some alarms can be reclassified. For example:
•The default severity level for Alarm A020 (calibration factors unentered) is Fault, but you can reconfigure it to either Informational or Ignore.
•The default severity level for Alarm A102 (drive
For a list of all status alarms and default severity levels, see Table
To configure alarm severity:
•Using ProLink II, see Figure
•Using a DeviceNet tool, refer to Table
a.Set the alarm index (Attribute 18).
b.Set the severity for that alarm (Attribute 19).
Note: This functionality is not available via the display menus.
Table 8-8 Status alarms and severity levels
|
| Default |
| Affected by |
Alarm code | ProLink II message | severity | Configurable? | fault timeout? |
|
|
|
|
|
A001 | (E)EPROM Checksum Error (CP) | Fault | No | No |
|
|
|
|
|
A002 | RAM Error (CP) | Fault | No | No |
|
|
|
|
|
A003 | Sensor Failure | Fault | Yes | Yes |
|
|
|
|
|
A004 | Temperature Sensor Failure | Fault | No | Yes |
|
|
|
|
|
A005 | Input Overrange | Fault | Yes | Yes |
|
|
|
|
|
A006 | Not Configured | Fault | Yes | No |
|
|
|
|
|
A008 | Density Overrange | Fault | Yes | Yes |
|
|
|
|
|
A009 | Transmitter Initializing/Warming Up | Ignore | Yes | No |
|
|
|
|
|
A010 | Calibration Failure | Fault | No | No |
|
|
|
|
|
A011 | Zero Too Low | Fault | Yes | No |
|
|
|
|
|
A012 | Zero Too High | Fault | Yes | No |
|
|
|
|
|
A013 | Zero Too Noisy | Fault | Yes | No |
|
|
|
|
|
A014 | Transmitter Failed | Fault | No | No |
|
|
|
|
|
A016 | Line RTD Temperature | Fault | Yes | Yes |
|
|
|
|
|
A017 | Meter RTD Temperature | Fault | Yes | Yes |
|
|
|
|
|
A020 | Calibration Factors Unentered (FlowCal) | Fault | Yes | No |
|
|
|
|
|
A021 | Incorrect Sensor Type (K1) | Fault | No | No |
|
|
|
|
|
A029 | PIC/Daughterboard Communication Failure | Fault | No | No |
|
|
|
|
|
A030 | Incorrect Board Type | Fault | No | No |
|
|
|
|
|
A031 | Low Power | Fault | No | No |
|
|
|
|
|
A032 | Meter Verification/Outputs In Fault | Fault | No | No |
|
|
|
|
|
A033 | Sensor OK, Tubes Stopped by Process | Fault | Yes | Yes |
|
|
|
|
|
A102 | Drive Overrange/Partially Full Tube | Info | Yes | No |
|
|
|
|
|
A104 | Calibration in Progress | Info | Yes(1) | No |
|
|
|
|
|
A105 | Slug Flow | Info | Yes | No |
|
|
|
|
|
A107 | Power Reset Occurred | Info | Yes | No |
|
|
|
|
|
66 | Micro Motion® Model 2400S Transmitters for DeviceNet™ |