Optional Configuration

Some alarms can be reclassified. For example:

The default severity level for Alarm A020 (calibration factors unentered) is Fault, but you can reconfigure it to either Informational or Ignore.

The default severity level for Alarm A102 (drive over-range) is Informational, but you can reconfigure it to either Ignore or Fault.

For a list of all status alarms and default severity levels, see Table 8-8.(For more information on status alarms, including possible causes and troubleshooting suggestions, see Table 11-2.)

To configure alarm severity:

Using ProLink II, see Figure B-3.

Using a DeviceNet tool, refer to Table C-7and:

a.Set the alarm index (Attribute 18).

b.Set the severity for that alarm (Attribute 19).

Note: This functionality is not available via the display menus.

Table 8-8Status alarms and severity levels

 

 

Default

 

Affected by

Alarm code

ProLink II message

severity

Configurable?

fault timeout?

 

 

 

 

 

A001

(E)EPROM Checksum Error (CP)

Fault

No

No

 

 

 

 

 

A002

RAM Error (CP)

Fault

No

No

 

 

 

 

 

A003

Sensor Failure

Fault

Yes

Yes

 

 

 

 

 

A004

Temperature Sensor Failure

Fault

No

Yes

 

 

 

 

 

A005

Input Overrange

Fault

Yes

Yes

 

 

 

 

 

A006

Not Configured

Fault

Yes

No

 

 

 

 

 

A008

Density Overrange

Fault

Yes

Yes

 

 

 

 

 

A009

Transmitter Initializing/Warming Up

Ignore

Yes

No

 

 

 

 

 

A010

Calibration Failure

Fault

No

No

 

 

 

 

 

A011

Zero Too Low

Fault

Yes

No

 

 

 

 

 

A012

Zero Too High

Fault

Yes

No

 

 

 

 

 

A013

Zero Too Noisy

Fault

Yes

No

 

 

 

 

 

A014

Transmitter Failed

Fault

No

No

 

 

 

 

 

A016

Line RTD Temperature Out-of-Range

Fault

Yes

Yes

 

 

 

 

 

A017

Meter RTD Temperature Out-of-Range

Fault

Yes

Yes

 

 

 

 

 

A020

Calibration Factors Unentered (FlowCal)

Fault

Yes

No

 

 

 

 

 

A021

Incorrect Sensor Type (K1)

Fault

No

No

 

 

 

 

 

A029

PIC/Daughterboard Communication Failure

Fault

No

No

 

 

 

 

 

A030

Incorrect Board Type

Fault

No

No

 

 

 

 

 

A031

Low Power

Fault

No

No

 

 

 

 

 

A032

Meter Verification/Outputs In Fault

Fault

No

No

 

 

 

 

 

A033

Sensor OK, Tubes Stopped by Process

Fault

Yes

Yes

 

 

 

 

 

A102

Drive Overrange/Partially Full Tube

Info

Yes

No

 

 

 

 

 

A104

Calibration in Progress

Info

Yes(1)

No

 

 

 

 

 

A105

Slug Flow

Info

Yes

No

 

 

 

 

 

A107

Power Reset Occurred

Info

Yes

No

 

 

 

 

 

66

Micro Motion® Model 2400S Transmitters for DeviceNet