Chapter 2 Command Summary

HP 53131A/132A Command Summary

Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)

Keyword/Syntax

Parameter Form

Std/

Comments

 

 

New

 

 

 

 

 

:DIAGnostic (Cont.)

 

 

 

:CALibration (Cont.)

 

 

 

:ROSCillator

 

New

Subtree. ROSCillator is an abbreviation for

 

 

 

Reference OSCillator.

:AUTO

ONCE OFF

New

ONCE calibrates the timebase. This command is

 

 

 

usable only if the instrument contains the medium or

 

 

 

high stability oscillator option.

:STATus?

 

New

Query only. Returns status of last calibration.

 

 

 

0 = pass; 1 = fail.

:TINTerval

 

New

Subtree.

:FINE

[1 2 3 4]

New

Event; no query. Four steps which calibrate out

 

 

 

Ch1, 2 electrical path length differences.

:QUICk

 

New

Event; no query. Calibrates out Ch1, 2 electrical path

 

 

 

length differences.

 

 

 

 

:DISPlay

 

Std

Subsystem. Controls the selection and presentation

 

 

 

of textual information on the display.

:ENABle

<Boolean>

Std

Controls whether the whole display is visible.

:MENU

 

Std

Subtree.

[:STATe]

OFF

Std

Sets the Counter to switch from the menu display to

 

 

 

the result display.

[:WINDow]

 

Std

Subtree.

:TEXT

 

Std

Subtree. Allows for the display of textual information.

:FEED

“[:]CALCulate2 ”

Std

Sets which data flow is fed into the display. “CALC2”

 

“[:]CALCulate3 ”

 

specifies the raw measurement, scaled/offset

 

 

 

measurement, or Limit Graph display. “CALC3”

 

 

 

specifies the statistics result display.

:RADix

COMMa DPOint

New

Sets the character used to separate integral and

 

 

 

fractional portions of a number. (USA numerical

 

 

 

convention is Decimal POint.)

 

 

 

 

:FETCh

 

Std

See Measurement Instructions in this table.

 

 

 

 

:FORMat

 

Std

Subsystem. Sets a data format for transferring

 

 

 

numeric information.

[:DATA]

ASCii REAL

Std

Sets the data format.

 

 

 

 

:HCOPy

 

New

 

:CONTinuous

<Boolean>

New

Enables or disables printing results.

 

 

 

 

2-28

Programming Guide