Chapter 4 Command Reference

Introduction

:DIAGnostic:CALibration:TINTerval:FINE[1234]

These event commands, when performed in the appropriate order and with the appropriate calibration signals supplied to both channels, calibrate out the differences in electrical path length between Channel 1 and Channel 2.

 

 

Before sending any of these commands, the appropriate calibration signal must be

 

 

supplied to both channels. Please see the section titled “To Initiate the Calibration

 

 

Routine” in Chapter 2 of the Operating Guide for details involved in providing the

 

 

calibration signals. The Operating Guide also discusses the difference between this

 

 

calibration and the alternate (simpler) choice, :DIAG:CAL:TINT:QUICk.

 

 

The command :DIAG:CAL:TINT:FINE1 invokes the first step of the calibration,

 

 

:DIAG:CAL:TINT:FINE2 invokes the second, and so on. Each step corresponds

 

 

to the calibration signals produced by each of the four numbered buttons on the

 

 

HP 59992A J06 Time Interval Calibrator.

 

This is a four-step calibration— all four steps must be successfully completed in

NOTE

 

 

 

sequence, starting with step one and ending with step four. To restart the

 

 

calibration (from step one), power cycle the Counter. A good precaution is to use

 

 

an HP-IB program (see the programming example titled “To Read and Store

 

 

Calibration Data” in Chapter 3 of the Programming Guide) to read and store the

 

 

calibration factors prior to initiating any calibration(s).

 

 

 

Comments

∙ No query.

 

 

∙ The calibration values are stored in non-volatile memory, and are unaffected

 

 

by power-on, save/recall, and *RST.

 

 

∙ Use :DIAG:CAL:STAT? to check for successful calibration, after all four

 

 

steps have been completed.

 

 

∙ Note, early versions of the Counter do not support these commands.

Related

Scale & Offset / POWER (Calibration Menu)

Front-Panel

 

Keys

 

Programming Guide

4-35