Chapter 4 Command Reference

:CALCulate3 Subsystem

Query Response Numeric data transferred as ASCII bytes in <NR3> format with eleven significant digits.

Comments

∙ *RST: 0.0000000000

∙ This command couples :CALC2:LIM:LOW to the same value.

∙ Updating the lower limit value causes the limit counts (:CALC2:LIM:FCO, :CALC2:LIM:PCO) to be cleared.

∙ The front panel menu item is not always able to display all of the significant digits of this value. When this is the case, the displayed value is different from the actual value in that the displayed value has been rounded. However, using the front panel Enter key, while this value is in the 11-digit display, will update the actual value to the displayed (rounded) value.

:CALCulate3:LFILter:STATe <Boolean>

Query Response

Comments

Related

Front-Panel

Keys

Sets or queries the statistics filter enable. When set to ON, only measurements (scaled and offset if math is enabled) which are within the filter limits are combined into the statistics processing. When set to OFF, all measurements, whether they are within or without the filter limits are combined into the statistics processing.

Single ASCII-encoded byte, 0 or 1.

A value of 0 indicates OFF; a value of 1 indicates ON. *RST: OFF

Stats

:CALCulate3:LFILter:UPPer[:DATA] <numeric_value> [HZ S DEG]

Sets or queries the statistics filter upper limit.

If limit filtering is enabled (:CALC3:LFIL:STAT ON), any measurements above this value will not be combined into the statistics computation.

If math is enabled (:CALC:MATH:STATe ON), the limit value specified should take into account that the filtering is on measurements that have been scaled and offset.

4-24

Programming Guide

Page 190
Image 190
HP 132A, 53131A manual CALCulate3LFILterSTATe Boolean, CALCulate3LFILterUPPerDATA numericvalue HZ S DEG