Chapter 3 Programming Your Universal Counter for Remote Operation

 

 

Elements of SCPI Commands

 

 

Response Message Data Types

 

 

Table 3-3 contains explanations of response data types.

 

 

Table 3-3. Response Message Data Types

 

 

 

 

 

Type

 

Description

 

 

 

 

<NR1>

 

This numeric representation has an implicit radix point.

 

 

 

 

 

 

 

 

 

 

 

 

The maximum number of characters in <NR1> response data is 17 (maximum

 

 

16 digits, 1 sign).

 

 

 

<NR2>

 

This numeric representation has an explicit radix point.

 

 

 

 

 

 

 

 

 

 

 

The maximum number of characters in <NR2> response data is 17 (maximum

 

 

15 mantissa digits, 1 sign, 1 decimal point).

 

 

 

<NR3>

 

This numeric representation has an explicit radix point and an exponent.

 

 

 

 

 

 

 

 

 

 

The maximum number of characters in <NR3> response data is 22 (maximum

 

 

15 mantissa digits, 2 signs, 1 decimal point , 1 ‘E’character, 3 exponent digits).

 

 

 

Not a Number

 

Not a Number is represented by the value 9.91E37 . (Not a Number is defined in

 

 

IEEE 754). The Counter responds with this numeric value when queried for a

 

 

floating point number it cannot provide. This value will be formatted as an

 

 

<NR3>.

 

 

 

 

 

Programming Guide

3-17

Page 85
Image 85
HP 53131A, 132A manual Response Message Data Types, Type Description