Chapter 3 Programming Your Universal Counter for Remote Operation

Elements of SCPI Commands

3a To set the address to “15”, perform the following:

a. Press s key.

HP-IB: 03 is displayed. Note that “0” digit appears and is highlighted, indicating that this digit will change when the

d or f arrow key is pressed.

b.Press d key.

HP-IB: 13 is displayed.

c.Press g key

HP-IB:13 is displayed, but now the “3” digit is highlighted, indicating

that this digit will change when the d or f arrow key is pressed.

 

 

 

d. Press d key twice or until 15 is displayed.

 

 

 

 

HP-IB:

15 is displayed.

 

 

 

e.

Go on to step 4.

3b

To set the HP-IB mode to “TALK,” perform the following:

 

 

 

a.

Press s

key.

 

 

 

 

HP-IB:

03 is displayed. Note that “0” digit appears and is highlighted,

 

 

 

 

indicating that this digit will change when the

 

 

 

 

d or

f

arrow key is pressed.

 

 

 

b.

Press d

key until “TALK” is displayed.

 

 

 

 

HP-IB:

TALK is displayed.

 

 

 

c. Go on to step 4.

4

 

 

Press Enter key.

 

BE SURE to press the Enter key to complete the entry.

NOTE

 

 

 

 

 

 

 

 

The address/mode is now stored in non-volatile memory, and does not change when power is cycled or after a remote interface reset.

Programming Guide

3-5