Chapter 2 Command Summary

HP 53131A/132A Command Summary

Table 2-2A. HP 53132A (S/N Prefix 3646 and Above) Time IntervalArming SCPI Command Summary

Keyword/Syntax

Parameter Form

Std/

Comments

 

 

New

 

 

 

 

 

[:SENSe] (Cont.)

 

 

 

:TINTerval

 

New

Subtree. Controls the time interval (including

 

 

 

Time Interval, Rise Time, Fall Time, Dutycycle,

 

 

 

and Pulse Width functions) measuring

 

 

 

capabilities of the instrument.

:ARM

 

New

Subtree. Synchronizes the time interval start and

:ESTART

 

New

stop arm with events.

 

 

:LAYer2

 

New

 

:SLOPe

POSitive NEGative

New

Sets the slope of the external start arm signal

 

 

 

used in external arming Time Interval

 

 

 

measurements (including Time Interval, Rise

 

 

 

Time, Fall Time, Dutycycle, and Pulse Width).

:SOURce

IMMediate EXTernal

New

Sets the start arm for time interval (including

 

 

 

Time Interval, Rise Time, Fall Time, Dutycycle,

[:LAYer[1]]

 

New

and Pulse Width) measurements.

 

 

:ECOunt

<numeric_value>

New

Sets the event count used to delay the start arm

 

 

 

for Time Interval measurements. Only applies

 

 

 

when [:SENS]:TINT:ARM:ESTART:SOUR INT1

 

 

 

is selected.

:SOURce

IMMediate TIMer

New

Sets the start arm delay for Time Interval

 

INTernal1

 

measurements.

:TIMer

<numeric_value>

New

Sets the time used to delay the start arm for

 

 

 

Time Interval measurements. Only applies when

 

 

 

[:SENS]:TINT:ARM:ESTART:SOUR TIM is

:ESTOP

 

New

selected.

 

 

:LAYer2

 

New

 

:SLOPe

POSitive NEGative

New

Sets the slope of the external stop arm signal

 

 

 

used in external arming Time Interval

 

 

 

measurements.

:SOURce

IMMediate EXTernal

New

Sets the stop arm for Time Interval

[:LAYer[1]]

 

New

measurements.

 

 

:ECOunt

<numeric_value>

New

Sets the event count used to delay the stop arm

 

 

 

for Time Interval measurements. Only applies

 

 

 

when [:SENS]:TINT:ARM:ESTOP:SOUR INT2 is

 

 

 

selected.

 

 

 

 

2-38