Chapter 4 Command Reference

:CALCulate2 Subsystem

Comments Query only.

:CALCulate2:LIMit:LOWer[:DATA] <numeric_value> [HZ S DEG]

Sets or queries the lower limit used for limit testing.

When the result is less than the lower limit, a fail is reported; when the result is equal to the lower limit, a fail is not reported.

If math is enabled (:CALC:MATH:STATe ON), the limit value specified should take into account that the limit testing is on measurements that have been scaled and offset.

<numeric_value> Range

<numeric_value> Resolution

Query Response

Comments

Related

Front-Panel

Keys

-9.9999990000E+12 to -1.0000000000E-13, 0.0000000000, +1.0000000000E-

13to +9.9999990000E+12.

11digits

Numeric data transferred as ASCII bytes in <NR3> format with eleven significant digits.

*RST: 0.0000000000

This command couples :CALC3:LFIL:LOW to the same value.

Updating the lower limit value causes the limit counts (:CALC2:LIM:FCO, :CALC2:LIM:PCO) to be cleared.

The front panel menu item is not always able to display all of the significant digits of this value. When this is the case, the displayed value is different from the actual value in that the displayed value has been rounded. However, using the front panel Enter key, while this value is in the 11-digit display, will update the actual value to the displayed (rounded) value.

Uppr & Lower

4-16

Programming Guide