<numeric_value> Range

Query Response

Chapter 4 Command Reference

Introduction

This command simulates the pressing or a front-panel key.

The <numeric_value> is a key code value.

This command puts an entry in the Key Queue (just as any front-panel key press does). The length of the Key Queue is 500.

The keys and their corresponding key codes are listed in the following table.

Key

Key Code

Key

Key Code

 

 

 

 

Freq & Ratio

1

d

25

 

 

 

 

Time & Period

4

f

28

 

 

 

 

Other Meas

2

s

26

 

 

 

 

Gate & ExtArm

5

g

27

 

 

 

 

Uppr & Lower

7

Channel 1 Trigger/Sensitivity

13

 

 

 

 

Limit Modes

8

Channel 1 50Ω /1MΩ

14

 

 

 

 

Scale & Offset

10

Channel 1 DC/AC

15

 

 

 

 

Stats

11

Channel 1 X10 Attenuate

16

 

 

 

 

Recall(Utility)

3

Channel 1 100kHz Filter

17

 

 

 

 

Save & Print

6

Channel 2 Trigger/Sensitivity

19

 

 

 

 

Run

9

Channel 2 50Ω /1MΩ

20

 

 

 

 

Stop/Single

12

Channel 2 DC/AC

21

 

 

 

 

+/-

29

Channel 2 X10 Attenuate

22

 

 

 

 

Enter

30

Channel 2 100kHz Filter

23

 

 

 

 

1 to 30, except 18 and 24

Numeric data transferred as ASCII bytes in <NR1> format.

The query returns the key code for the last key pressed. A value of -1 indicates the queue (last in, first out) is empty.

Programming Guide