Chapter 4 Command Reference

:DISPlay Subsystem

:DISPlay[:WINDow]:TEXT:FEED “[:]CALCulate2” “[:]CALCulate3”

Query Response

Comments

Sets or queries what data flow is fed into the display.

Choose from the following <data_handle> strings:

“[:]CALCulate2”— should be used to direct any result other than the statistics to the result display

Specifically this would select one of the following results for the result display:

raw measurement (if math is disabled— :CALC:MATH:STAT OFF)

the scaled/offset measurement (if math is enabled— :CALC:MATH:STAT ON)

the limit graph (if limit testing is enabled with the graphic display— :CALC2:LIM:STAT ON, :CALC2:LIM:DISP GRAPh)

“[:]CALCulate3”— should be used to direct the statistical result

(if statistics are enabled, :CALC3:AVER[:STAT] ON) to the result display; the particular statistic displayed is determined by :CALC3:AVER:TYPE.

A string is returned: “CALC2” or “CALC3.”

*RST: “:CALCulate2”

Refer to the section titled “How to Program the Counter to Display Results” in Chapter 3 of this guide.

Related Stats

Front-Panel

Keys

4-38

Programming Guide

Page 204
Image 204