Chapter 4 Command Reference

Introduction

[:SENSe]:TOTalize:ARM[:STARt]:SOURce IMMediate EXTernal

Sets or queries the start arm for Totalize measurements.

Query Response A sequence of ASCII-encoded bytes: IMM or EXT

Comments ∙ *RST: IMMediate

When both start and stop arm are set to IMMediate, use :INIT[:IMM] to begin totalizing and :ABORt to terminate. No valid totalize result will exist until the “auto” totalize measurement is terminated.

Related Gate & ExtArm

Front-Panel

Keys

[:SENSe]:TOTalize:ARM:STOP:SLOPe POSitive NEGative

Sets or queries the slope of the external stop arm signal used in external arming Totalize measurements.

Query Response A sequence of ASCII-encoded bytes: POS or NEG

Comments ∙ *RST: NEGative

Only applies when [:SENS]:TOT:ARM:STOP:SOUR EXT is selected.

Related Gate & ExtArm

Front-Panel

Keys

Programming Guide

4-105