Chapter 4 Command Reference

Introduction

Related Stats

Front-Panel

keys

:CALCulate3:AVERage:COUNt:CURRent?

Queries the current count (that is, the number of data values collected for statistical computation).

Query Response ∙ Numeric data transferred as ASCII bytes in <NR1> format.

Range is 0 to 1,000,000.

If :CALC3:AVER[:STATe] is OFF, error -221 is generated.

If the current measurement is Totalize or Voltage Peaks, 0 is returned and error -221 is generated.

Comments ∙ Query only.

No statistics results exist until the :CALC3:AVER:COUN:CURR? is equal to the specified :CALC3:AVER:COUN.

Related Stats

Front-Panel

keys

:CALCulate3:AVERage[:STATe] <Boolean>

Sets or queries the statistics post-processing enable .

Query Response ∙ Single ASCII-encoded byte, 0 or 1.

A value of 0 indicates OFF; a value of 1 indicates ON.

Comments ∙ *RST: OFF

When this enable is ON, and :TRIG:COUN:AUTO is ON, and [:SENS]:FUNC[:ON] is not Totalize or Voltage Peak function, then :INIT[:IMM] initiates a complete block of measurements. See :TRIG:COUN:AUTO in this chapter for specifics.

When this enable is OFF, :INIT[:IMM] always initiates a single measurement.

Programming Guide