Chapter 5 Errors

Error Types

Table 5-2. Errors (Continued)

Number

Error String

Cause

 

 

 

 

 

 

 

-400

Query error

This is the generic query error.

 

 

-410

Query INTERRUPTED

Indicates that a condition causing an INTERRUPTED Query

 

 

error occurred. For example, a query followed by DAB or GET

 

 

before a response was completely sent.

 

 

-420

Query UNTERMINATED

Indicates that a condition causing an UNTERMINATED Query

 

 

error occurred. For example, the Counter was addressed to talk

 

 

and an incomplete program message was received.

 

-430

Query DEADLOCKED

Indicates that a condition causing a DEADLOCKED Query error

 

 

occurred. For example, both input buffer and output buffer are

 

 

full and the Counter cannot continue.

 

 

-440

Query UNTERMINATED

Indicates that a query was received in the same program

 

after indefinite response

message after a query requesting an indefinite response

 

 

( for example, *IDN? or *OPT?) was executed.

 

+2000

Offset calibration on channel 1 failed

:DIAGnostic:CALibration:INP:OFFS:AUTO ONCE failed.

+2001

Offset calibration on channel 2 failed

:DIAGnostic:CALibration:INP2:OFFS:AUTO

ONCE

failed.

+2002

Gain calibration on channel 1 failed

:DIAGnostic:CALibration:INP:GAIN:AUTO ONCE failed.

+2003

Gain calibration on channel 2 failed

:DIAGnostic:CALibration:INP2:GAIN:AUTO

ONCE

failed.

+2004

Interpolator calibration failed

:DIAGnostic:CALibration:INTerpolator:AUTO

ONCE, *CAL?,

 

 

or :CALibration[:ALL]? failed.

 

 

+2005

Oscillator calibration failed

:DIAGnostic:CALibration:ROSCillator:AUTO

ONCE

failed.

+2006

Time interval calibration failed

:DIAGnostic:CALibration:TINTerval:QUICk failed.

 

+2007

Measurement hardware calibration

A measurement calibration failed on the last measurement.

 

failed

 

 

 

+2008

Measurement interpolator calibration

Interpolator calibration failed on the last measurement;

 

failed

therefore, no valid measurement was taken.

 

 

+2009

No valid external timebase

You have selected external reference and there is no external

 

 

reference applied to the rear-panel Ref In connector, or the

 

 

external signal is not an allowed frequency.

 

 

+2010

External timebase failed during

You have selected the auto reference mode and the Counter

 

measurement

detected that the external reference became invalid

during the

 

 

measurement. Therefore, the current result is not valid, and the

+2011

Recall setup failed; hardware failure

Counter switches to using the internal reference.

 

*RCL failed.

 

 

 

Recall setup failed; empty register

*RCL failed because the register specified for recall is empty.

+2012

Save setup failed

*SAV failed, or *RCL couldn ’t save to register 0.

 

+2013

EEPROM failed

A request to update a setting which is stored in the EEPROM

 

 

(:SYST:COMM:SER:TRAN:BAUD,

 

 

 

 

:SYST:COMM:SER:TRAN:PAR,

 

 

 

 

:SYST:COMM:SER:TRAN:PACE,

 

 

 

 

:SYST:COMM:SER:CONT:DTR, :DISP[:WIND]:TEXT:RAD,

 

 

:CAL[:DATA], or :DIAG:CAL: ...) resulted in a hardware failure.

 

 

 

 

 