Chapter 4 Command Reference

:CALCulate[1] Subsystem

:CALCulate[1]:FEED “[:]SENSe[1]”

Sets or queries the data flow to be fed into the CALCulate[1] block.

Since the Counter can only sense one function at a time, there is only one valid parameter.

Query Response The string “SENS” is returned.

Comments *RST: “SENSe[1]”

:CALCulate[1]:IMMediate

Query Response

Comments

This command is an event that causes the Counter to recalculate existing data without re-acquiring data. (This recalculation also happens automatically when any change is made to the

:CALCulate[12] subsystems while :CALC:IMM:AUTO is ON.)

:CALC:IMM? is semantically equivalent to :CALC:IMM;DATA?. The query form outputs the results of the new calculation.

This command will not affect:

:CALC2:LIM:FCO

:CALC2:LIM:PCO :CALC3: ...

Result will be formatted according to :FORMat[:DATA] ASCii REAL setting.

When ASCii format is used, numeric data is transferred as ASCII bytes in <NR3> format. The number of significant digits will range from 1 to 15, depending on the measurement resolution.

Only significant digits will be returned.

If no valid result exists, Not a Number 9.91E37 is returned and error -230 is generated.

If the current measurement is Totalize or Voltage Peaks, Not a Number 9.91E37 is returned and error -221 is generated.

This command causes post-processing to occur in the :CALCulate2 subsystem, as well as the CALCulate subsystem.