Chapter 2 Command Summary

Front Panel to SCPI Command Maps

Table 2-3. HP 53131A/132A *RST State (Continued)

Command Header

Parameter

State

 

 

 

*EMC

<NRf>

0 (i.e., disabled)

:FORMat[:DATA]

ASCii REAL

ASCii

 

 

 

:HCOPy:CONTinuous

<Boolean>

OFF

 

 

 

:INITiate:AUTO

<Boolean>

OFF

:INITiate:CONTinuous

<Boolean>

OFF

 

 

 

:INPut[12]:ATTenuation

1 10

1

:INPut[12]:COUPling

AC DC

AC

:INPut[12]:FILTer[:LPASs]:STATe

<Boolean>

OFF

:INPut[12]:IMPedance

<numeric_value> [OHM]

1E6 OHM

 

 

 

[:SENSe]:EVENt[12]:HYSTeresis:RELative

<numeric_value> [PCT]

100 PCT *

[:SENSe]:EVENt[12]:LEVel[:ABSolute]:AUT

<Boolean>

ON

O

<numeric_value> [PCT]

50 PCT

[:SENSe]:EVENt[12]:LEVel:RELative

POSitive NEGative

POSitive

[:SENSe]:EVENt[12]:LEVel:SLOPe

“INPut1 INPut2”

“INPut2”

[:SENSe]:EVENt2:FEED

 

(i.e., separate mode)

 

POSitive NEGative

POSitive

[:SENSe]:FREQuency:ARM[:STARt]:SLOPe

IMMediate EXTernal

IMMediate

[:SENSe]:FREQuency:ARM[:STARt]:SOURce

<numeric_value>

4

 

[:SENSe]:FREQuency:ARM:STOP:DIGits

POSitive NEGative

NEGative

[:SENSe]:FREQuency:ARM:STOP:SLOPe

IMMediate EXTernal TIMer DIGits

TIMer

[:SENSe]:FREQuency:ARM:STOP:SOURce

<numeric _value> [S]

100E3 S

[:SENSe]:FREQuency:ARM:STOP:TIMer

ON

ON

[:SENSe]:FREQuency:EXPected[123]:AUTO

<sensor_function>

 

[:SENSe]:FUNCtion[:ON]

“FREQuency 1”

POSitive NEGative

 

 

POSitive

[:SENSe]:PHASe:ARM[:STARt]:SLOPe

IMMediate EXTernal

IMMediate

[:SENSe]:PHASe:ARM[:STARt]:SOURce

ON OFF ONCE

 

 

ON

[:SENSe]:ROSCillator:EXTernal:CHECk

<Boolean>

ON

[:SENSe]:ROSCillator:SOURce:AUTO

 

 

 

 

 

________________________________

*For prior firmware revisions (3317, 3335, and 3402), *RST sets up “100%”meaning maximum sensitivity.

For current firmware revision, *RST sets up “0%”meaning minimum hysteresis.

Programming Guide

2-41