HP 53131A, 132A Keyword/Syntax Parameter Form Std Comments New INITiate, INPut12, INPut3, MEASure

Models: 132A 53131A

1 330
Download 330 pages 55.08 Kb
Page 53
Image 53

Chapter 2 Command Summary

Front Panel to SCPI Command Maps

Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)

Keyword/Syntax

Parameter Form

Std/

Comments

 

 

New

 

 

 

 

 

:INITiate

 

Std

Subsystem. Controls the initiation of measurements.

:AUTO

<Boolean>

New

AUTO ON enables the Counter to automatically stop

 

 

 

measuring on a limit test failure. AUTO OFF

 

 

 

disables the automatic stop.

:CONTinuous

<Boolean>

Std

Sets the enable for continuously initiated

 

 

 

measurements.

[:IMMediate]

 

Std

Event; no query. Causes the instrument to initiate

 

 

 

the number of measurements specified by

 

 

 

:TRIGger:COUNt:AUTO.

 

 

 

 

:INPut[12]

 

Std

Subsystem. Controls the characteristics

 

 

 

of the instrument ’s input ports. :INPut1= channel 1

 

 

 

and :INPut2= channel 2

:ATTenuation

1 10

Std

Sets input attenuation.

:COUPling

AC DC

Std

Sets input coupling.

:FILTer

 

Std

Subtree. Allows a low pass filter to be inserted in the

 

 

 

path of the measurement signal.

[:LPASs]

 

Std

Subtree. Controls the Low PASs filter.

[:STATe]

<Boolean>

Std

Sets the Low PASs filter enable.

:FREQuency?

 

Std

Query only. Returns the cutoff frequency of the low

 

<numeric_value>

Std

pass filter. Units are Hertz.

:IMPedance

Sets input impedance (50 Ω or 1 M Ω ).

 

[OHM]

 

 

 

 

 

 

:INPut3

 

Std

Subsystem. Queries the characteristics of the

 

 

 

Counter’s input channel 3.

:COUPling?

 

Std

Query only. Returns channel 3 input coupling.

:IMPedance?

 

Std

Query only. Returns channel 3 input impedance.

 

 

 

 

:MEASure

 

Std

See Measurement Instructions in this table.

 

 

 

 

Programming Guide

2-29

Page 53
Image 53
HP 53131A, 132A manual Keyword/Syntax Parameter Form Std Comments New INITiate, INPut12, INPut3, MEASure