Query Response

Comments

Chapter 4 Command Reference

[:SENSe] Subsystem

The string “<function> <channel>[,<channel>] ” is returned.

The string omits default nodes (XNONe) and uses short form mnemonics. If the channel specifier(s) are set to default value(s), no channel specifier is returned in response. If the channel specifier(s) are not set to default value(s), they will be returned in the response with a single space separating the first channel specifier from the function name.

For example:

“FREQ” would be returned for frequency on Channel 1.

“FREQ 2” would be returned for frequency on Channel 2.

“FREQ:RAT” would be returned for frequency ratio of Channel 1 to Channel 2.

“FREQ:RAT 1,3” would be returned for frequency ratio of Channel 1 to Channel 3.

*RST: “FREQuency 1”

If the optional channel specification is omitted from the <sensor_function>, a default channel selection is made. For Frequency, Period, Rise Time, Fall Time, Pulse Width, Duty Cycle, Totalize, and Voltage Peaks, the default is Channel 1. For Time Interval, Frequency Ratio, and Phase, the default is Channel 1 to Channel 2.

When the sensor function is changed to [:][XNONe:]TINTerval [1,2], the input routing is set to separate ([:SENS]:EVEN2:FEED “INP2”). Use [:SENS]:EVEN2:FEED “INP1” to select common-mode Time Interval.

When the sensor function is changed to Rise Time, the Channel 1 trigger level is set to 10%, the Channel 2 trigger level is set to 90%, and auto-triggering is enabled.

When the sensor function is changed to Fall Time, the Channel 1 trigger level is set to 90%, the Channel 2 trigger level is set to 10%, and auto-triggering is enabled.

When the sensor function is changed to Duty Cycle or Pulse Width, the Channel 1 trigger level is set to 50%, the Channel 2 trigger level is set to 50%, and auto-triggering is enabled.

4-90

Programming Guide