Chapter 4 Command Reference

Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)

Descriptions of the Measurement Functions— <function> (Cont.)

:MEASure[:SCALar][:VOLTage]:TOTalize:TIMed? [<gate_time>][,(@1)]

Measures Totalize during the specified <gate time>.

The totalize gate mode is set to “time,” and the totalize gate time is defaulted or set to the specified value.

The trigger settings are coupled so that the measurement channel has auto-trigger disabled with a positive slope.

<gate_time>

 

 

range:

For short gate time:

100E-5 to 999E-5 seconds

 

For long gate time: 10E-3 to 1000.000 seconds

resolution:

For short gate time:

1E-5 seconds

 

For long gate time: 1E-3 seconds

default:

.100 second

 

<source_list>

 

 

range:

(@1)

 

default:

(@1)

 