Chapter 2 Command Summary

Front Panel to SCPI Command Maps

Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)

Keyword/Syntax

Parameter Form

Std/

Comments

 

 

New

 

 

 

 

 

:CALCulate3 (Cont.)

 

New

Subtree. Limit FILter for statistics.

:LFILter

 

:LOWer

 

New

Subtree.

[:DATA]

<numeric_value> [HZ S

New

Sets the statistics filter lower limit.

 

DEG]

 

 

:STATe

<Boolean>

New

Sets the statistics filter enable.

:UPPer

 

New

Subtree.

[:DATA]

<numeric_value> [HZ S

New

Sets the statistics filter upper limit.

:PATH?

DEG]

Std

Query only. Returns LFIL, AVER.

 

 

 

 

 

:CALibration

 

Std

Subsystem.

[:ALL]?

 

Std

Query only. Causes an internal interpolator

 

 

 

self-calibration.

:COUNt?

 

New

Query only. Returns value indicating number

 

 

 

of times the Counter has been calibrated.

:DATA

<arbitrary block>

Std

Transfers the calibration data (input gain,

 

 

 

input offset, reference oscillator, and time

:SECurity

 

New

interval).

:CODE

<NRf>

New

No query. Sets the calibration security code.

:STATe

<Boolean>, <NRf>

New

Enables or prevents calibration of the

 

 

 

Counter. Query returns security status. 0 =

 

 

 

unsecure; calibration allowed. 1 = secure;

 

 

 

calibration disallowed.

 

 

 

 

:CONFigure

 

Std

See Measurement Instructions in this table.

 

 

 

 

:DIAGnostic

 

Std

Subsystem.

:CALibration

 

New

Subtree.

:INPut[12]

 

New

Subtree. 1 2 specifies channel.

:GAIN

 

New

Subtree.

:AUTO

ONCE OFF

New

ONCE calibrates input gain.

:OFFSet

 

New

Subtree.

:AUTO

ONCE OFF

New

ONCE calibrates input offset.

:INTerpolator

 

New

Subtree.

:AUTO

ONCE OFF ON

New

ONCE calibrates the interpolators.

 

 

 

 

Programming Guide

2-27

Page 51
Image 51
HP 53131A, 132A manual Keyword/Syntax Parameter Form Std Comments New CALCulate3, CALibration, CONFigure, DIAGnostic