Chapter 3 Programming Your Universal Counter for Remote Operation

Programming Examples

To Measure the Statistics of 50 Measurements (HP BASIC)

10! This program instructs the counter to take 50 period measurements.

20! The counter is put into SINGLE measurement mode. The number of

30! measurements to take is programmed using ":CALC3:AVER:COUNT 50"

40! The counter is told to stop after 50 measurements using

50! ":TRIG:COUNT:AUTO ON"

60! At the end of 50 measurements, the statistics are calculated and

70! sent to the computer.

80! ASCII format is used to preserve resolution.

90! In this example, the status reporting structure is used to alert

100! the program that the statistics are ready.

110! The "*OPC" command and the "*ESE 1 " command are used together

120! to generate an output from the Event Status Register when

130! the measurement is complete. The output of this register is

140! used as an input to the Service Request Register. In order for the

150! Service Request Register to be able to use that input, the "*SRE 32"

160! command must be used. This enables the Service Request Register to

170! assert the SRQ line when the measurement is complete.

180! Note, that the *OPC command must be sent prior to every measurement

190! in order to enable the OPC bit. See Line # 520.

200INTEGER I,Num_meas

210DIM Sdev$[22],Mean$[22],Minimum$[22],Maximum$[22]

220

Num_meas=50

! Statistics based on Num_meas measurements

230

ASSIGN @Count TO 703

 

240

CLEAR 703

! Clear the counter and interface

250

OUTPUT @Count;"*RST"

! Reset the counter

260

OUTPUT @Count;"*CLS"

! Clear event registers and error queue

270OUTPUT @Count;"*SRE 0 " ! Clear service request enable register

280OUTPUT @Count;"*ESE 0 " ! Clear event status enable register

290OUTPUT @Count;":STAT:PRES" ! Preset enable registers and transition

300

! filters for operation and questionable

310

! status structures.

320

OUTPUT @Count;":FUNC 'PER 1'" ! Measure Period on channel 1.

330

! Note that the functions must be

340

! a quoted string. The actual

350

! string sent to the counter

360

! is 'PER 1'.

370OUTPUT @Count;":FREQ:ARM:STAR:SOUR IMM" ! These three lines enable

380OUTPUT @Count;":FREQ:ARM:STOP:SOUR TIM" ! time arming with a 0.01

390OUTPUT @Count;":FREQ:ARM:STOP:TIM .01" ! second gate time.

400OUTPUT @Count;":DISP:TEXT:FEED 'CALC3'" ! Display statistics

410OUTPUT @Count;":CALC3:AVER:TYPE SDEV" !Display the standard deviation

420

OUTPUT @Count;":CALC3:AVER ON"

! Enable statistics

430

OUTPUT @Count;":CALC3:AVER:COUNT ";Num_meas ! Do statistics on num_meas

440

 

 

! measurements.

450

OUTPUT @Count;":TRIG:COUNT:AUTO ON "

!Take Num_meas measurements

460

OUTPUT @Count;"*ESE 1"

! "*ESE 1" is used so bit 5

470

 

!of the service request register will allow

480OUTPUT @Count;"*SRE 32" ! an SRQ when measurement complete.

490ON INTR 7 GOTO Get_averages ! Goto Get_averages on interrupt.

500

ENABLE INTR 7;2

! Enable interrupt on counter SRQ.

510

PRINT "Waiting for measurement to complete"

3-64