Chapter 4 Command Reference

:DIAGnostic Subsystem

:DIAGnostic:CALibration:TINTerval:QUICk

This event command calibrates out the differences in electrical path length between channels 1 and 2.

Before sending this command, connect to channel 1 a square wave of approximate frequency 10 MHz, but more importantly with a rapid rise time.

Comments

Related

Front-Panel

Keys

No query.

The calibration values are stored in non-volatile memory, and are unaffected by power-on, save/recall, and *RST.

Use DIAG:CAL:STAT? to check for successful calibration.

Scale and Offset/POWER (Calibration menu).

4-36

Programming Guide