Chapter 4 Command Reference

:INITiate Subsystem

When a single measurement is in progress (:INIT:CONT is OFF):

Error -213 (Init ignored) is generated and the state of INIT:CONT is unaffected by :INIT:CONT ON.

Error -210 (Trigger error) is generated by INIT:CONT OFF.

Note that the Counter powers up with :INIT:CONT set to ON, but *RST sets :INIT:CONT to OFF.

Related Run

Front-Panel

Keys

:INITiate[:IMMediate]

This event command causes the instrument to initiate either a single measurement or a block of measurements.

When

:TRIG:COUN:AUTO is OFF, or :CALC3:AVER[:STAT] is OFF, or [:SENS]:FUNC[:ON] is Totalize function, or [:SENS]:FUNC[:ON] is any Voltage Peaks function,

then :INIT[:IMM] initiates a single measurement.

When

:TRIG:COUN:AUTO is ON, and :CALC3:AVER[:STAT] is ON, and [:SENS]:FUNC[:ON] is not Totalize or Voltage Peak function,

then :INIT[:IMM] initiates a complete block of measurements. See :TRIG:COUN:AUTO for specifics.

4-46

Programming Guide