Chapter 4 Command Reference

*OPC

(Operation Complete Command)

*OPC

(Operation Complete Command)

*OPC

This event command enables the OPC bit (bit 0) in the Standard Event Status Register to be set upon the transition of the measurement cycle from measuring to idle. (Note — For HP 53131A/132A with serial number prefix 3646 and above

If the measurement cycle is already idle, this command will immediately set the OPC bit.) See the section titled “Standard Event Status Register Group,” page 3- 24, in Chapter 3 of this guide for a detailed description of the Standard Event Status Register’s Operation Complete bit.

This event command is “disabled” by *CLS, *RST, Device Clear

(page 4-31), power-on, or upon the transition of the measurement cycle from measuring to idle.

This event command has no query form.

See the section titled “Using the *OPC Command to Assert SRQ,” page 3-48, in Chapter 3 for an example using this command.

4-132

Programming Guide

Page 298
Image 298
HP 132A, 53131A manual Operation Complete Command, Opc