Chapter 4 Command Reference

Introduction

[:SENSe]:EVENt2:FEED “[:]INPut[1] [:]INPut2”

Query Response

Comments

Sets or queries the common/separate enable.

Feeding the :INPut2 subsystem to the [:SENSe]:EVENt2 subsystem corresponds to separate; feeding the :INPut1 subsystem to the [:SENSe]:EVENt2 subsystem corresponds to common.

The string “INP” or “INP2” is returned.

*RST: “INPut2”

Only the Time Interval function allows either the separate or common setting. All of the other functions have a fixed input route setting.

Related Trigger/Sensitivity

Front-Panel

Keys

[:SENSe]:EVENt[12]:HYSTeresis:RELative <numeric_value> [PCT]

The operation of this command is determined, in part, by the firmware revision of the Counter.

Current Firmware Revision

Sets or queries the size of the hysteresis window as a percentage of the allowable hysteresis. For example, 0% is the minimum hysteresis setting and 100% is the maximum hysteresis setting.

Specifying 100% or MAXimum provides the greatest noise immunity (lowest sensitivity), while specifying 0% or MINimum provides the least noise immunity (most sensitive).

Prior Firmware Revisions: 3317, 3335, and 3402

Sets or queries the sensitivity of the input channel as a percentage of the allowable sensitivity. For example, 0% is the minimum sensitivity setting and 100% is the maximum sensitivity setting.

Specifying 100% or MAXimum provides the most sensitivity

(least noise immunity), while specifying 0% or MINimum provides the lowest sensitivity (greatest noise immunity).

Programming Guide

4-81

Page 247