HP 132A, 53131A manual SENSeTOTalize Subtree, SENSeTOTalizeARMSTARtSLOPe POSitive NEGative

Models: 132A 53131A

1 330
Download 330 pages 55.08 Kb
Page 270
Image 270

Chapter 4 Command Reference

[:SENSe] Subsystem

[:SENSe]:TOTalize Subtree

This subtree controls the Totalize measuring capabilities of the instrument.

[:SENSe]:TOTalize:ARM Subtree

This subtree is used to synchronize the Totalize start and stop arm with events.

The following combination of start/stop arming sources are valid:

STARt:SOURce

STOP:SOURce

Front-Panel Gating Settings

 

 

GATE

STOP

 

 

 

 

IMMediate

IMMediate

AUTO

_______

 

 

 

 

IMMediate

TIMer

TIME

_______

 

 

 

 

EXTernal

EXTernal

EXTERNAL

POS, NEG

 

 

 

 

EXTernal

TIMer

EXTERNAL

TIME

 

 

 

 

[:SENSe]:TOTalize:ARM[:STARt]:SLOPe POSitive NEGative

Sets or queries the slope of the external start arm signal used in external arming

Totalize measurements.

Query Response A sequence of ASCII-encoded bytes: POS or NEG

Comments ∙ *RST: POSitive

Only applies when [:SENS]:TOT:ARM[:STAR]:SOUR EXT is selected.

Related

Front-PanelGate & ExtArm

Keys

4-104

Programming Guide

Page 270
Image 270
HP 132A, 53131A manual SENSeTOTalize Subtree, SENSeTOTalizeARMSTARtSLOPe POSitive NEGative, SENSeTOTalizeARM Subtree