Chapter 4 Command Reference

[:SENSe] Subsystem

[:SENSe]:TOTalize:ARM:STOP:SOURce IMMediate EXTernal TIMer

Sets or queries the stop arm for Totalize measurements.

Query Response A sequence of ASCII-encoded bytes: IMM, EXT or TIM

Comments ∙ *RST: TIMer

When both start and stop arm are set to IMMediate, use :INIT[:IMM] to begin totalizing and :ABORt to terminate. No valid totalize result will exist until the “auto” totalize measurement is terminated.

Related Gate & ExtArm

Front-Panel

Keys

[:SENSe]:TOTalize:ARM:STOP:TIMer <numeric_value> [S]

<numeric_value> Range

<numeric_value Resolution

Query Response

Comments

Related

Front-Panel

Keys

Sets or queries the gate time used in arming Totalize measurements.

For short gate time: 1E-3 to 99.99E-3 seconds

For long gate time: 10 0E-3 to 1000.000 seconds

For short gate time: 0.01E-3 seconds

For long gate time: 1E-3 seconds

Numeric data transferred as ASCII bytes in <NR3> format with six significant digits.

*RST: 100E-3 S

Only applies when [:SENS]:TOT:ARM:STOP:SOUR TIM is selected. Gate & ExtArm

4-106

Programming Guide

Page 272
Image 272