Chapter 4 Command Reference

Introduction

Descriptions of the Measurement Functions— <function> (Cont.)

:MEASure[:SCALar][:VOLTage]:NWIDth? [<reference>][,(@1)]

Measures Negative Pulse Width.

The measurement arming is coupled to “auto.”

The <reference> specifies the point on the pulse where the negative pulse width is determined. This point can be specified as either a percentage or an absolute voltage. The default units are percent.

The <reference> parameter is used to configure the trigger settings. When the <reference> is specified as a percentage (or defaulted),

auto-trigger is enabled and auto-trigger levels are set to the specified (or defaulted) reference value. When the <reference> is specified in volts, auto-trigger is disabled and the absolute trigger levels are set to the specified reference value.

<reference>

 

percent range:

0 to 100 [PCT]

percent resolution:

10%

voltage range:

For volts if X1 Attenuation: -5.125V to +5.125V

 

For volts if X10 Attenuation: -51.25V to +51.25V

voltage resolution:

For volts if X1 Attenuation: .005V

 

For volts if X10 Attenuation: .05V

default:

50 PCT

<source_list>

 

range:

(@1)

default:

(@1)

Programming Guide

4-67