Chapter 4 Command Reference

Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)

Descriptions of the Measurement Functions— <function> (Cont.)

:MEASure[:SCALar][:VOLTage]:RISE:TIME? (Cont.)

<lower_reference>

percent range:

0 to 100 [PCT]

percent resolution:

10%

voltage range:

For volts if X1 Attenuation: -5.125V to +5.125V

 

For volts if X10 Attenuation: -51.25V to +51.25V

voltage resolution:

For volts if X1 Attenuation: .005V

 

For volts if X10 Attenuation: .05V

default:

10 PCT

<upper_reference>

 

percent range:

0 to 100 [PCT]

percent resolution:

10%

voltage range:

For volts if X1 Attenuation: -5.125V to +5.125V

 

For volts if X10 Attenuation: -51.25V to +51.25V

voltage resolution:

For volts if X1 Attenuation: .005V

 

For volts if X10 Attenuation: .05V

default:

90 PCT

<source_list>

 

range:

(@1)

default:

(@1)

4-72

Programming Guide

Page 238
Image 238