Chapter 4 Command Reference

Introduction

*RST

(Reset Command)

*RST

This event command performs an instrument reset .

The reset performs the following:

sets instrument settings to their *RST states,

disables macros,

places instrument in “Operation Complete Idle State” and “Operation Complete Query Idle State,” and

clears (empties) the Key Queue

The reset does not affect:

the macros defined with *DMC,

the calibration data,

the Service Register Enable or the Standard Event Status Enable,

the Output Queue, and

the IEEE 488.1 address or the state of the IEEE 488.1 interface.

See the section titled “*RST Response,” page 2-36, in Chapter 2 of this guide for a complete listing of the *RST state.

Each command description in this chapter (Chapter 4) includes the *RST state in the “Comment” portion of the definition.

Programming Guide

4-137