Chapter 4 Command Reference

:CALCulate3 Subsystem

If the current measurement is Totalize or Voltage Peaks, Not a Number 9.91E37 is returned and error -221 is generated.

Comments ∙ Query only.

The last calculated result remains valid until a new computation is made or a relevant instrument state is modified.

Related Stats

Front-Panel

keys

:CALCulate3:AVERage:CLEar

This command is an event that causes the Counter to:

invalidate the statistics results,

clear the statistics current count to 0, and

report the negative status condition (NOT Computing Statistics) to bit 8 of the Operation Status Register.

Comments If :CALC3:AVER[:STATe] is OFF, error -221 is generated.

:CALCulate3:AVERage:COUNt <numeric_value>

Sets or queries the number of measurements to combine for statistics processing .

After :CALC3:AVER:COUNt measurements is reached, a new set of :CALC3:AVER:COUNt measurements must be acquired before another statistics computation will occur.

<numeric_value> 2 to 1,000,000

Range

<numeric_value> 1

Resolution

Query Response Numeric data transferred as ASCII bytes in <NR1> format.

Comments *RST: 100