Chapter 2 Command Summary

Front Panel to SCPI Command Maps

Table 2-2A. HP 53132A (S/N Prefix 3646 and Above) Time Interval

Arming SCPI Command Summary (Continued)

Keyword/Syntax

Parameter Form

Std/

Comments

 

 

New

 

 

 

 

 

[:SENSe] (Cont.)

 

 

 

:TINTerval (Cont.)

 

 

 

:ARM (Cont.)

 

 

 

:ESTOP (Cont.)

 

 

 

[:LAYer[1]]

IMMediate TIMer

New

Sets the stop arm delay for Time Interval

:SOURce

 

INTernal2

 

measurements.

:TIMer

<numeric_value>

New

Sets the time used to delay the stop arm for

 

 

 

Time Interval measurements. Only applies

 

 

 

when [:SENS]:TINT:ARM:ESTOP:SOUR TIM is

[:STARt]

 

New

selected.

 

 

:SLOPe

POSitive NEGative

New

Sets the slope of the external start arm signal

 

 

 

used in external arming Tme Interval

 

 

 

measurements (including Time Interval, Rise

 

 

 

Time, Fall Time, Dutycycle, and Pulse Width).

 

 

 

Only applies when

 

 

 

[:SENS]:TINT:ARM[:STAR]:SOUR EXT is

 

 

 

selected.

:SOURce

IMMediate EXTernal

New

Sets the start arm for time interval (including

 

 

 

Time Interval, Rise Time, Fall Time, Dutycycle,

 

 

 

and Pulse Width) measurements.

:STOP

 

New

Subtree.

:SOURce

IMMediate TIMer

New

Sets the stop arm (i.e., the delay) for Time

 

 

 

Interval measurements.

:TIMer

<numeric_value>

New

Sets the time used to delay the stop arm for

 

 

 

Time Interval measurements. Only applies

 

 

 

when [:SENS]:TINT:ARM:STOP:SOUR TIM is

 

 

 

selected.

 

 

 

 

Programming Guide

2-39