Chapter 4 Command Reference

Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)

Descriptions of the Measurement Functions— <function> (Cont.)

:MEASure[:SCALar][:VOLTage]:FREQuency? (Cont.)

default:

value which indicates 4 digits of resolution for the

 

specified <expected_value>

Ch3 <expected_value>

 

range:

100 MHz to 3.00 GHz

resolution:

<expected_value> should be within 10% of input

 

frequency for optimum arming configuration

default:

500 MHz

Ch3 <resolution>

 

description:

value indicates decade corresponding to least significant

 

digit of the result

range:

1E-7 to1E7 Hz

 

value which indicates 3 to 15 digits of resolution for the

 

specified <expected value>

resolution:

<resolution> should use a mantissa of 1.0 and be an

 

even power of 10

default:

value which indicates 4 digits of resolution for the

 

specified <expected_value>

<source_list>

range:

(@1) (@2) (@3)

default:(@1)

4-64

Programming Guide

Page 230
Image 230
HP 132A, 53131A manual MEASureSCALarVOLTageFREQuency?