Chapter 2 Command Summary

HP 53131A/132A Command Summary

Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)

Keyword/Syntax

Parameter Form

 

 

 

 

Std/

Comments

 

 

 

 

 

 

New

 

 

 

 

 

 

 

 

 

[:SENSe] (Cont.)

 

 

 

 

 

Std

Subtree. Selects the <sensor function> to

:FUNCtion

 

 

 

 

 

 

 

 

 

 

 

 

be sensed by the instrument.

[:ON]

<sensor_function> (See below)

 

Std

Sets the <sensor function> to be sensed by

 

 

 

 

 

 

 

the instrument.

 

“[:][XNONe:]DCYCle

[1] ”

 

New

(Positive) duty cycle on channel 1.

 

“[:][XNONe:]PDUTycycle

[1] ”

 

New

(Positive) duty cycle on channel 1.

 

“[:][XNONe:]FALL:TIME

[1] ”

 

New

Fall Time on channel 1.

 

“[:][XNONe:]FTIMe [1] ”

 

 

New

Fall Time on channel 1.

 

“[:][XNONe:]FREQuency

[1 2 3] ”

 

Std

Frequency on channel 1, 2, or 3.

 

“[:][XNONe:]FREQuency:RATio

 

Std

1Frequency Ratio 1 to 2, 1 to 3, 2 to 1, or 3

 

[1,2 1,3 2,1 3,1] ”

 

 

to 1.

 

“[:][XNONe:]NWIDth

[1] ”

 

New

Negative Pulse Width on channel 1.

 

“[:][XNONe:]PERiod

[1 2 3] ”

 

Std

Period on channel 1, 2, or 3.

 

“[:][XNONe:]PHASe

[1,2] ”

 

Std

Phase of channel 1 relative to channel 2.

 

“[:][XNONe:]PWIDth

[1] ”

 

New

Positive Pulse Width on channel 1.

 

“[:][XNONe:]RISE:TIME [1] ”

 

New

Rise Time on channel 1.

 

“[:][XNONe:]RTIMe [1] ”

 

 

New

Rise Time on channel 1.

 

“[:][XNONe:]TINTerval

[1,2] ”

 

New

Time Interval channel 1 to channel 2.

 

“[:][XNONe:]TOTalize

[1] ”

 

Std

Totalize on channel 1.

 

“[:][XNONe:]VOLTage:MAXimum [1 2] ”

 

Std

Voltage Maximum on channel 1 or 2.

 

“[:][XNONe:]VOLTage:MINimum [1 2] ”

 

New

Voltage Minimum on channel 1 or 2.

 

“[:]{XNONe:]VOLTage:PTPeak [1 2]

 

New

Voltage Peak to Peak on Channel 1 or 2.

:PHASe

 

 

 

 

 

New

Subtree. Controls the phase measuring

 

 

 

 

 

 

New

capabilities of the instrument.

:ARM

 

 

 

 

 

Subtree.

[:STARt]

 

 

 

 

 

New

Subtree. Synchronizes Phase start arm with

 

POSitive NEGative

 

 

 

 

New

events.

:SLOPe

 

 

 

 

Sets the slope of the external start arm

 

 

 

 

 

 

 

signal used in external arming phase

 

 

 

 

 

 

 

measurements. Only applies when

 

 

 

 

 

 

 

[:SENS]:PHAS:ARM[:STAR]:SOUR EXT is

 

IMMediate EXTernal

 

 

 

 

New

selected.

:SOURce

 

 

 

 

Sets the start arm for phase

:ROSCillator

 

 

 

 

 

Std

measurements.

:EXTernal

 

 

 

 

 

Std

Subtree. Controls the Reference

:CHECk

ON OFF ONCE

 

 

 

 

New

OSCillator.

 

 

 

 

 

 

Std

Subtree.

:FREQuency?

 

 

 

 

 

Set the enable for checking the validity and

 

 

 

 

 

 

 

presence of the external reference.

 

 

 

 

 

 

 

Query only. Returns the frequency value of

 

 

 

 

 

 

 

the external reference oscillator.

 

 

 

 

 

 

______________________________

 

 

 

1For HP 53131A firmware revision s below 3335, only Ratio 1 to 2 and Ratio 1 to 3 were offered.

2-34