Chapter 4 Command Reference

Introduction

The [:SENS]:TINT:ARM:EST ART[:LAYer[1]]:ECO command has no affect on the following measurements: Risetime, Falltime, Duty Cycle, and Pulse Width.

[:SENSe]:TINTerval:ARM:ESTART[:LAYer[1]]:SOURce IMMediate TIMer INTernal[1]

Sets or queries the start arm delay for Time Interval measurements.

Query Response

A sequence of ASCII-encoded bytes: IMM or TIM or INT[1]

 

Comments

∙ *RST: IMMediate

 

 

∙ The [:SENS]:TINT:ARM:ESTART[:LAYer[1]]:SOUR command has no

 

affect on the following measurements: Risetime, Falltime, Duty Cycle, and

 

Pulse Width.

Related

Gate & ExtArm

 

Front-Panel

 

Keys

 

[:SENSe]:TINTerval:ARM:ESTART[:LAYer[1]]:TIMer <numeric_value> [S]

Sets or queries the time used to delay the start arm for Time Interval measurements.

<numeric_value>

100E-9 to 0.999,999,9 seconds

Range

 

<numeric_value>

100E-9 seconds

Resolution

 

Query Response

Numeric data transferred as ASCII bytes in <NR1> format.

 

Comments

∙ *RST: 100E-9 seconds

 

Related

Front-Panel

Keys

Only applies when [:SENS]:TINT:ARM:ESTART[:LAYer[1]]:SOUR TIM is selected.

The [:SENS]:TINT:ARM:ESTART[:LAYer[1]]:TIM command has no affect on the following measurements: Risetime, Falltime, Duty Cycle, and Pulse Width.

Gate & ExtArm

Programming Guide

4-101

Page 267
Image 267
HP 53131A, 132A manual SENSeTINTervalARMESTARTLAYer1TIMer numericvalue S, Related Front-Panel Keys