Chapter 4 Command Reference

Introduction

:DIAGnostic:CALibration:INPut[12]:OFFSet:

AUTO ONCE OFF

Query Response

Comments

Related

Front-Panel

Keys

Calibrates the channel 1 or 2 input trigger OFFSet when the ONCE parameter is used.

Before sending this command, BE SURE to disconnect any input signal from the appropriate input.

A sequence of ASCII-encoded bytes: OFF

The calibration values are stored in non-volatile memory, and are unaffected by power-on, save/recall, and *RST.

Use :DIAG:CAL:STAT? to check for successful calibration.

After calibration is completed, the state of this command’s parameter is OFF. Scale and Offset/POWER (Calibration menu).

:DIAGnostic:CALibration:INTerpolator:AUTO ONCE OFF ON

Calibrates the interpolator circuit in the Counter when the ONCE parameter is used.

AUTO ON enables automatic interpolator calibration on every measurement. AUTO OFF disables automatic interpolator calibration.

Query Response

Comments

A sequence of ASCII-encoded bytes: OFF or ON

*RST: ON

Use :DIAG:CAL:STAT? to check for successful calibration.

After ONCE calibration is completed, the state of this command’s parameter is OFF.

When :DIAG:CAL:INT is set to OFF, the Counter reports the positive status condition (questionable Time, Frequency, and Phase) to bits 2, 5, and 6 of the Questionable Status Register. When :DIAG:CAL:INT is set to ON, the Counter reports the negative status condition (NOT questionable Time, Frequency, and Phase) to bits 2, 5, and 6 of the Questionable Status Register.

This enable is unaffected by save/recall.

Programming Guide

4-33