Chapter 4 Command Reference

Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)

Descriptions of the Measurement Functions— <function> (Cont.)

:MEASure[:SCALar][:VOLTage]:PERiod? [<expected_value>[,<resolution>]][, (@1)(@2)(@3)]

Measures Period.

The measurement arming mode is set to “digits.” The Counter uses the <expected_value> and <resolution> parameters to configure the number of digits of resolution arming setting.

The Channel 1 or Channel 2 trigger settings are coupled so that the measurement channel has auto-trigger enabled at 50% with a positive slope.

Ch1, Ch2 <expected_value>

range:

4.4 ns to 10.0 sec

resolution:

<expected value> should be within 10% of input period

 

for optimum arming configuration

default:

100 ns

Ch1, Ch2 <resolution>

 

description:

value indicates decade corresponding to least significant

 

digit of the result

range:

1E-23 to 1E-2 sec

 

value which indicates 3 to 15 digits of resolution for the

 

specified <expected value>

resolution:

<resolution> should use a mantissa of 1.0 and be an even

 

power of 10

default:

value which indicates 4 digits of resolution for the

 

specified <expected_value>

4-68

Programming Guide

Page 234
Image 234