Chapter 4 Command Reference

:DIAGnostic Subsystem

:DIAGnostic Subsystem

This subsystem controls the remote calibration of the Counter.

All of the calibration values, with the exception of the interpolator values, are stored in non-volatile memory and are unaffected by

power-on, save/recall, and *RST.

Any of the commands which perform a calibration, with the exception of the interpolator calibration, will generate error -221 if the user tries to execute a calibration while the Counter is secured. (Note, this will not occur in early revisions of the Counter because calibration security does not exist.) Please refer to the :CALibration:SECurity subtree for command specifics regarding calibration security.

:DIAGnostic:CALibration:INPut[12]:GAIN:

AUTO ONCE OFF

Query Response

Comments

Related

Front-Panel

Keys

Calibrates the channel 1 or 2 input trigger GAIN when the ONCE parameter is used.

Before sending this command, connect a +5V source to the appropriate input.

A sequence of ASCII-encoded bytes: OFF

The calibration values are stored in non-volatile memory, and are unaffected by power-on, save/recall, and *RST.

Use :DIAG:CAL:STAT? to check for successful calibration.

After calibration is completed, the state of this command’s parameter is OFF. Scale and Offset/POWER (Calibration menu).

4-32

Programming Guide

Page 198
Image 198
HP 132A, 53131A manual DIAGnostic Subsystem, DIAGnosticCALibrationINPut12GAIN, Auto Once OFF