Reference Manual

00809-0100-4530, Rev BA July 2009

Rosemount 5300 Series

DIAGNOSTICS DEVICE ERRORS

Table F-17. Device Errors Diagnostics

In addition to the BLOCK_ERR and XD_ERROR parameters, more detailed information on the measurement status can be obtained via DIAGN_DEV_ALERT. Table F-17lists the potential errors and the possible corrective actions for the given values. The corrective actions are in order of increasing system level compromises. The first step should always be to reset the gauge and then if the error persists, try the steps in Table F-17.Start with the first corrective action and then try the second.

Bit Number

Value of

Description

Corrective action

DIAGN_DEV_ALERT

 

 

 

 

0

No alarm active

 

0

0x00000001

Reserved

 

1

0x00000002

FF card to gauge comm fault

Replace gauge

2

0x00000004

Level Measurement Failure

Verify device installation and configuration

3

0x00000008

Temperature Measurement

Check ambient temperature. If ambient

 

 

Failure

temperature ok replace the device.

4

0x00000010

Volume measurement failure

Check volume configuration

5

0x00000020

Database Error

Load default database to the device and

 

 

 

reconfigure the device

6

0x00000040

HW Error

Replace device

7

0x00000080

Microwave Unit Error

Replace device

8

0x00000100

Configuration Error

Load default database to the device and

 

 

 

reconfigure the device

9

0x00000200

SW Error

Replace device

10

0x00000400

Invalid Strap Table

Check the Strapping Table configuration

11

0x00000800

Internal Temp Warning

Check ambient temperature at installation site

12

0x00001000

Database Warning

Check the device configuration

13

0x00002000

HW Warning

Check device installation and configuration

14

0x00004000

Microwave Unit Warning

Check device installation and configuration

15

0x00008000

Configuration Warning

Check the device configuration

16

0x00010000

SW Warning

Check device installation and configuration

17

0x00020000

Simulation Mode

Use Simulation Method under advanced

 

 

 

configuration tool to get the device out of

 

 

 

Simulation Mode

18

0x00040000

Volume Range Warning

Verify Strapping Table

19

0x00080000

Software Write Protected

 

20

0x00100000

Vapor Compensation Not

 

 

 

Calibrated

 

21

0x00200000

Vapor DC Estimation Limited

 

22

0x00400000

Reference Reflector Not

 

 

 

Found

 

23

0x00800000

BaseLine too low

 

24

0x01000000

BaseLine too high

 

25

0x02000000

Gain control

 

26

0x04000000

Unsupported Combination of

 

 

 

Functions

 

27

0x08000000

 

 

28

0x10000000

 

 

29

0x20000000

Probe Missing

Check probe connection

30

0x40000000

Interface Measurement

Check interface measurement configuration

 

 

Failure

 

31

0x80000000

Hardware Write Protected

 

F-9