Emerson Process Management 5300 manual This product meets the Dual Seal Requirements of ANSI/ISA

Models: 5300

1 324
Download 324 pages 46.88 Kb
Page 236
Image 236

Rosemount 5300 Series

Reference Manual

00809-0100-4530, Rev BA July 2009

CANADIAN STANDARDS ASSOCIATION (CSA) APPROVAL

Intrinsic Safety

This product meets the Dual Seal Requirements of ANSI/ISA 12.27.01-2003.

Dual seal annunciation

The breach of the secondary seal is annunciated via product leakage from the antenna’s vents. The leakage will be visible and/or audible from the transmitter head’s threads.

Dual seal maintenance

No maintenance required. Verify proper operation by keeping leak path free from ice or contamination.

Cert. no. 1514653.

Figure B-14. Approval Label Canadian Standards Association (CSA) HART model

I6 Intrinsically Safe Ex ia.

Class I, Division 1, Groups A, B, C and D. Temperature code T4.

4-20 mA/HART model: Ui=30 V, Ii=130 mA, Pi=1.0 W, Ci=7.26 nF, Li=0 H.

Installation Drawing: 9240 030-937.

Ambient temperature limits: -50 °C to + 70 °C.

B-14

Page 236
Image 236
Emerson Process Management 5300 manual This product meets the Dual Seal Requirements of ANSI/ISA, Dual seal annunciation