Emerson Process Management 5300 manual Application Example, Radar Level Transmitter, Level Value

Models: 5300

1 324
Download 324 pages 46.88 Kb
Page 114
Image 114

Rosemount 5300 Series

Reference Manual

00809-0100-4530, Rev BA July 2009

Application Example 1

Radar Level Transmitter, Level Value

A level transmitter is measuring the level in a 33 ft. (10 m) high tank.

Figure 5-12. Situation Diagram

100%

33ft

(10 m)

0%

Solution

Table 5-9lists the appropriate configuration settings, and Figure 5-13illustrates the correct function block configuration.

Table 5-9. Analog Input

Function Block Configuration for

a typical Level Transmitter

Parameter

Configured Values

 

 

 

 

L_TYPE

Direct

 

XD_SCALE

Not Used

 

OUT_SCALE

Not Used

 

CHANNEL

CH1: Level

Figure 5-13. Analog Input

Function Block Diagram for a Level Measurement typical Level Transmitter

AI Function Block

 

OUT_D

 

 

 

 

OUT

To Another

 

 

 

Function Block

 

 

 

 

 

 

 

5-42

Page 114
Image 114
Emerson Process Management 5300 manual Application Example, Radar Level Transmitter, Level Value, Solution