REMOVING THE TRANSMITTER HEAD

FLANGE VERSION

Nut

 

Put the protection

Process Seal

 

plug here!

 

THREADED VERSION

Put the protection

Nut

Process Seal

plug here!

 

 

Adapter

1.Loosen the nut that connects the transmitter housing to the Process Seal.

2. Carefully lift the transmitter head.

3.Make sure the upper surface of the Process Seal is clean and the spring-loaded pin at the center of the Process Seal is properly inserted (the pin should move back when pushed into the hole).

4.Attach the protection plug to the Process Seal.

NOTE!

Do not remove the Process Seal from the adapter!