Table 3-9. Hole requirements on probes for mounting a centering disc.

Mounting a Centering Disc on Rigid Single Probes

The mounting of a centering disc on a rigid single lead probe requires one hole (8 mm probe) or two holes (13 mm probe) at certain distances from the end of the lead. The drilling fixture included in your shipment should be used to make the hole(s) according to Figure 3-14.

Probe

Minimum distance to hole

Number of

Hole diameter

 

from probe end

holes

 

8 mm

5 mm

1

3.5 mm

 

 

 

 

13 mm

7 mm (first hole)

2

3.5 mm

 

 

 

 

Figure 3-14. Use drill template for proper holes.

8 mm

13 mm

Rigid Single Lead probe (13 mm)

Bushing

Washer

Probe

Drilling fixtures

1.Mount the centering disc at the end of the probe.

2.Fasten the disc by inserting the split pins through the bushings and the probe.

Split pin

Split pin

3. Adjust distance by shifting hole for split pin in lower lockring.

Lower lockring

4. Secure the split pin.

3-26

Page 56
Image 56