Reference Manual

00809-0100-4530, Rev BA July 2009

Rosemount 5300 Series

A short weight for the single flexible 0.16 in. (4 mm) SST probe can be used for measuring close to the probe end. The height is 2 in. (50 mm) and the diameter is 1.5 in.

(37.5 mm). Option code W2.

For hot applications, the chamber should always be insulated to prevent personal injuries and to reduce the amount of energy needed for heating. See Figure 3-6. It is often an advantage, and sometimes even required, for the radar measurement:

In hot applications, insulation reduces the amount of condensation, since it prevents the upper part of the chamber from becoming a cold spot

Insulation prevents product solidification inside the chamber, and clogging of the inlet-pipes

Figure 3-6. Insulated Chamber.

See page 3-14for more information.

When mounting in a Rosemount 9901 chamber, the probe length to use can be calculated with these formulas:

Side-and-Side dimension: Probe length = Centre-to-Centre dimension + 19 in. (48 cm)

Side-and-Bottom dimension: Probe length = Centre-to-Centre dimension + 4 in. (10 cm)

Side-and-Side Side-and-Bottom

dimensiondimension

Centre-to-Centre

Centre-to-Centre

3-9

Page 39
Image 39
Emerson Process Management manual Rosemount 5300 Series