Reference Manual

00809-0100-4530, Rev BA July 2009

Rosemount 5300 Series

To calibrate Dynamic Vapor Compensation:

1.Either choose Device Specific Configuration in the Guided Setup (if the configuration is recommended) or Setup, Advanced.

2.Select the Vapor Compensation tab.

Figure C-16. Calibrate Dynamic

Vapor Compensation.

3.Click the Calibrate Vapor Compensation button.

4.Make sure Use Vapor Compensation is selected.

5.The echo from the reference reflector end can be viewed in the Echo Curve in RRM. Probes shorter than 6.6 ft. (2 m) will have the reflector end at 0.75 ft. (0.23 m) distance and probes longer than 6.6 ft. (2 m) will have the reflector end at 1.6 ft. (0.5 m).

6.Cycle power to finish.

In 375 Field Communicator, the calibration can be done by HART command [2, 7, 3].