Emerson Process Management 5300 manual Surface Pulse Not Found on

Models: 5300

1 324
Download 324 pages 46.88 Kb
Page 164
Image 164

Rosemount 5300 Series

Reference Manual

00809-0100-4530, Rev BA July 2009

Symptom

 

Possible cause

 

Action

The reference pulse is not detected.

The tank is full

Check the product level

 

The transmitter is configured with

Check that correct probe type is configured

 

 

wrong probe type

• Check Reference Amplitude Threshold

 

Reference Amplitude Threshold is

 

 

 

 

incorrect

 

 

 

 

 

 

 

Incorrect Interface Level reading.

Interface threshold incorrect

Adjust Interface threshold, see “Interface

 

Upper Product Dielectric constant

 

Pulse not Found” on page 7-8

 

 

incorrect

Check the Upper Product Dielectric

 

 

 

 

constant, see “Basic Configuration

 

 

 

 

Parameters” on page 5-4

Incorrect level reading.

Configuration error

Check the Tank Height parameter

 

Disturbing objects in the tank

Check status information and diagnostic

 

 

 

 

information

 

 

 

Check that the transmitter has not locked

 

 

 

 

on an interfering object, see “Disturbance

 

 

 

 

Echo Handling” on page 7-7

 

 

 

• Adjust the Surface amplitude threshold, see

 

 

 

 

“Surface Pulse Not Found” on page 7-5

 

 

 

 

 

Integral display does not work.

 

 

Check the display configuration

 

 

 

• Check loop power

 

 

 

Check Display connection

 

 

 

• Contact Emerson Process Management

 

 

 

 

Service Department(1)

FOUNDATION Fieldbus Card to

 

 

Verify Device Mode setting, should be

Transmitter Communication Fault

 

 

 

FOUNDATION Fieldbus

 

 

 

 

(Parameter: ENV_DEVICE_MODE)

 

 

 

• Restart method from Resource Block

 

 

 

• Reboot gauge (Cycle Power)

 

 

 

 

Level Measurement Failure

 

 

• Check Power Supply

 

 

 

Check the gauge configuration (Transducer

 

 

 

 

Block)

 

 

 

Check that the mechanical installation is

 

 

 

 

correct

Temperature Measurement Failure

 

 

• Check ambient temperature(2)

 

 

 

Restart gauge

 

 

 

• Contact Emerson Process Management

 

 

 

 

Service Department

 

 

 

 

Volume Measurement Failure

 

 

• Restart gauge

 

 

 

• Check gauge configuration using PC Based

 

 

 

 

configuration tool

No surface echo

 

 

Check signal strength

 

 

 

Restart transmitter

 

 

 

• See “Surface Pulse Not Found” on

 

 

 

 

page 7-5

 

 

 

 

 

DB Error/ Microwave Unit Error/

 

 

Restart transmitter

Configuration Error/ Other Error

 

 

Set database to default; load default

 

 

 

 

Database

 

 

 

Download Application Software

 

 

 

Call Service Center

SW Error/ Display Error/ Analog Out

 

 

Restart transmitter

Error

 

 

Call Service Center

 

 

 

 

 

(1)A malfunctioning display panel may only be replaced by service personnel at the Emerson Process Management Service Department. A display must not be replaced when the transmitter is in operation.

(2)If the 5300 transmitter has been exposed to temperatures outside the specified limits, the device may stop its normal operation.

7-34

Page 164
Image 164
Emerson Process Management 5300 manual Surface Pulse Not Found on