Reset DC Estimation: Resets DC estimation to the configured initial value,

 

forcing the device to start over estimating the product DC.

 

Use Static Product DC: Check this setting if you do not want the device to

 

estimate the product DC. This will force the device to use the configured initial

 

product DC.

ECHO TRACKING

Measurement with the Rosemount 5300 is based on the fact that the radar

 

signal pulses are reflected by the product surface. Different parameters are

 

used to track the measurement signal to achieve a reliable and stable

 

measurement. Normally, the echo tracking parameters are automatically set

 

by the 5300 transmitter, and no manual settings are needed. However, due to

 

the properties of the product, in rare cases it may be necessary to adjust the

 

echo tracking settings for optimum measurement performance. RRM

 

supports echo tracking settings in the Advanced Configuration window,

 

Echo Tracking tab.

Figure C-10. Echo Tracking in

 

RRM.

 

Automatic echo tracking settings are enabled by default. In the Advanced

Configuration window the Echo Tracking settings Echo Timeout and Close

Distance can be also be set manually.