Rosemount 5300 Series

Reference Manual

00809-0100-4530, Rev BA July 2009

Figure B-10. Approval Labels

Factory Mutual (FM) Foundation

Fieldbus model

I5 Intrinsically Safe for Class I, II, III, Division 1, Groups A, B, C, D, E, F and G.

Class I, Zone 0, AEX ia IIC T4 when installed per Control Drawing: 9240030-936.

Non-incendive for Class I, Division 2, Groups A, B, C and D. Suitable for Class II, III, Division 2, Groups F and G;

FOUNDATION Fieldbus model: Ui=30 Vdc, Ii=300 mA, Pi=1.3 W, Ci=0 nF, Li=0 H. Max operation 32 V, 25 mA.

Temperature code T4.

Ambient Temperature Limits: -50 °C to + 60 °C

Figure B-11. Approval Labels Factory Mutual (FM) FISCO model

IE Intrinsically Safe for Class I, II, III, Division 1, Groups A, B, C, D, E, F and G.

Class I, Zone 0, AEX ia IIC T4 when installed per Control Drawing: 9240030-936.

FISCO model: Ui=17.5 Vdc, Ii=380 mA, Pi=5.32 W, Li=Ci=0.

Temperature code T4.

Ambient Temperature Limits: -50 °C to + 60 °C

B-12