Reference Manual

00809-0100-4530, Rev BA July 2009

Rosemount 5300 Series

NOTE!

Note that for firmware version 2.A2 the Trim Near Zone is not possible in combination with Narrow Nozzles, as defined below:

50 mm< Nozzle height < 300 mm

Nozzle diameter < 2 in. for all single probes except 13 mm single rigid Nozzle diameter < 3 in. for 13 mm single rigid

Figure C-3. Trim Near Zone using RRM.

1.Click Trim Near Zone.

2.Select Trim near zone and click OK.

3.Click OK in the appearing dialog.

4.Wait 1 minute.

5.Restart the transmitter.

C-5

Page 253
Image 253
Emerson Process Management 5300 manual Figure C-3. Trim Near Zone using RRM, Click Trim Near Zone