To cut a coaxial probe:

L>49 in. (1250 mm)

Centering piece

NOTE!

 

The HTHP coaxial probe must not

 

be cut in field.

 

 

 

1. Insert the centering piece.

 

(The centering piece is

 

delivered from factory and

 

should be used to prevent the

 

spacers centering the rod from

 

coming loose).

 

2. Cut the tube to the desired length.

 

3. Move the centering piece.

 

4. Cut the rod inside the tube. Make

 

sure that the rod is fixed with the

 

centering piece while cutting.

 

• Pipes longer than

 

49 in. (1250 mm) can be

 

shortened by as much as

 

23.6 in. (600 mm)

 

• Pipes shorter than

 

49 in. (1250 mm) can be cut as

 

long as the remaining length is not

 

less than

 

15.7 in. (400 mm)

Maximum shortening 23.6 in. (600 mm)

L49 in. (1250 mm)

Minimum probe length 15.7 in. (400 mm)

5.Update the transmitter configuration to the new probe length, see “Tank and Probe Geometry“ on page 5-4and “Probe“ on page 5-21.