Rosemount 5300 Series

Reference Manual

00809-0100-4530, Rev BA July 2009

LCD ERROR MESSAGES

Figure 7-24. The 5300 Display Panel displaying an error message.

Table 7-11. Error messages displayed on the 5300 Display Panel.

Error Message

Error Message

Description

 

An error in the gauge data memory (RAM) has been

RAM FAIL

detected during the startup tests. Note: this resets the

 

gauge automatically.

 

 

FPROM FAIL

An error in the gauge program memory (FPROM) has

been detected during the startup tests. Note: this

 

resets the gauge automatically.

 

An error in the transmitter configuration memory

 

(EEPROM) has been detected. The error is either a

HREG FAIL

checksum error that can be solved by loading the

default database, or a hardware error.

 

 

NOTE: the default values are used until the problem is

 

solved.

 

 

OMEM FAIL

 

MWM FAIL

An error in the microwave module.

 

 

DPLY FAIL

An error in the LCD.

MODEM FAIL

Modem hardware failure.

 

 

AOUT FAIL

An error in the Analog Out Module.

OHW FAIL

An unspecified hardware error has been detected.

 

 

ITEMP FAIL

An error in the internal temperature measurement.

MEAS FAIL

A serious measurement error has been detected.

 

 

 

At least one configuration parameter is outside the

CONFIG FAIL

allowed range.

NOTE: the default values are used until the problem is

 

 

solved.

SW FAIL

An error has been detected in the transmitter software.

 

 

For more information on errors, see “Errors” on page 7-36.

7-42

Page 172
Image 172
Emerson Process Management 5300 manual LCD Error Messages, For more information on errors, see Errors on